Technical Specificatiosn

sensor:

OPUS fingerprobe diameter = 89 mm
length = 330 - 3500 mm
for pipes from DN 10, vessels,...
measuring range: particle size
solids concentration
media
0.01 µm - 3000 µm
31 classes, xmax/xmin < 1000
1 - 70 Vol % typ.
suspensions, emulsions
principle: ultrasound 100 kHz to 200 MHz
dynamics fixed: 150 dB
precision: 0.1dB typ.
evaluation: extinction function  library > 300 products
operation: process control
software
optional via PLC
WINDOWS XP Prof / 2000
performance: repeatability

comparability 
x10, x50, x90
measuring time
SD. < 0.5% (repeated sample)
SD. < 1.0% (riffled sample)
SD. < 5% mean rel. SD

< 1 minute, typ.
environment: temperature (product)
pressure
protection class
materials in contact
with media
≤120°C (standard) or 150° C (as option)
≤ 40 bar
IP65 or ATEX
adaptable to application
stainless steel, Sigradur®, Teflon®,

more...

OPUS
Particle size analysis with ultrasound 0.01 µm - 3000 µm
in suspensions and emusions of high concentration

symbol: wet dispersion for suspensions or emulsions
Add this page to my favorites Send this page to my colleague Print this page