Technical Specifications

sample coupler: customer screw sampler, etc ....
principle: fully adaptable to cutomer's existing sampler highest flexibility
analysis system: MYTOS 0.25 µm - 3500 µm
dispersion
principle:
RODOS-aerosol jet particle-particle-collisions
particle-wall-collisions
velocity gradients
analysis
principle:
HELOS laser diffraction λ = 632.8 nm
measurement: multi-element-detector
auto-alignment
measuring frequency
31 semi-circular elements
permanent
2000/s
evaluation: Fraunhofer MIE as an option
measuring range: optical modules R2 - R7
operation: at-line

RS232 / RS484-interface
optional via PLC
Windows 2000
performance: repeatability
comparability 
x10, x50, x90
SD. < 0.3 % (riffled sample)
SD. < 1 % mean rel. SD
/Δx/ < 2.5 % rel. max. SD.
environment: temperature (product)
pressure
protection class
optional
materials in contact
with media
sample disposal
max. 100°C
max. 10 bar
IP65
Ex
Stainless Steel or customised

back into process or external
QS-system: certification
validation

 

standard test procedure
FDA approved
Add this page to my favorites Send this page to my colleague Print this page
at-line Laser Diffraction with
MYTOS & VIBRI

at-line: MYTOS & VIBRI 

MYTOS & VIBRI at-line

MYTOS & VIBRI

Dimensions of Systems

The following downloads display examples of the dimensions of some combinations of MYTOS and VIBRI for at-line applications as pdf-files.

Complete or partial duplication of these documents is prohibited without written permission. All rights to the documentation are reserved by Sympatec GmbH.

No guarantee is assumed for correctness of the contents.

MYTOS & VIBRI at-line:

Symbol: Sympatec Cactus