Technical Specifications

sample coupler: TWISTER pipes 35 - 800 mm
principle: representative
sampling
scanning of complete pipe
cross-section on a spiral path
analysis system: MYTOS 0.25 µm - 3500 µm
dispersion
principle:
RODOS-aerosol jet particle-particle-collisions
particle-wall-collisions
velocity gradients
analysis
principle:
HELOS laser diffraction λ = 632.8 nm1
measurement: multi-element-detector
auto-alignment
measuring frequency
31 semi-circular elements
permanent
2000/s
evaluation: Fraunhofer MIE as an option
measuring range: optical modules R2 - R7
operation: on-line

RS 232 / RS484-interface
optional via PLC
Windows 2000
performance: repeatability
comparability 
x10, x50, x90
SD. < 0.3 % (riffled sample)
SD. < 1 % mean rel. SD
/Δx/ < 2.5 % rel. max. SD.
environment: temperature (product)
pressure bursts
protection class
optional
materials in contact
with media
sample disposal
max. 120°C (TWISTER), max. 60°C (MYTOS)
max. 10 bar
IP65
Ex, GMP
Stainless Steel or customised

back into process or external
QS-system: certification
validation

 

standard test procedure
FDA approved
on-line Laser Diffraction with
MYTOS & TWISTER
Add this page to my favorites Send this page to my colleague Print this page

on-line: MYTOS & TWISTER 

MYTOS & TWISTER on-line

MYTOS & TWISTER 100

Dimensions of Systems

The following downloads display examples of the dimensions of some combinations of MYTOS and TWISTER for in-line applications as pdf-files.

Complete or partial duplication of these documents is prohibited without written permission. All rights to the documentation are reserved by Sympatec GmbH.

No guarantee is assumed for correctness of the contents.

MYTOS & TWISTER on-line:

Symbol: Sympatec Cactus