Symbol: Sympatec Cactussymbol: wet dispersion for suspensions or emulsions

INHALER side view with discus holder, 90° elbow, pre-separator, central stage, fine particle collector and wash bottle

Aerosol dispersion of MDIs, DPIs and Nebulizers for particle size analysis with the INHALER
mounted on the bench of a HELOS-VARIO/KF

Technical specifications and dimensions.

The INHALER is a versatile adapter for the investigation and the development of all kinds of inhalers, MDI's, propellant gas sprays or pump sprays.

The central stage with the measuring chamber and the vacuum control unit can be adapted to the various needs by a flexible tool kit comprising

  • a discus adapter - for the adaptation of discus shaped inhalers
  • elongation front cylinder - expand the distance from the mouth piece to the measuring volume
  • 90° elbow - simulated the human throat, customer specific designs can integrated easily
  • pre-separator - removes coarse particles from the measuring chamber
  • Venturi meter - for precise measurement of the air flow rate
  • fine particle collector - collects the particles with highest efficiency at high throughputs and lowest pressure drop for subsequent (chemical) analysis.
  • wash bottle - as with fine particle collector, but for medium throughputs
  • vacuum pump - for the extraction

Air flow rate and inclination angle (for nebulizers) are controlled by software. Ambient temperature, humidity, air flow rate (via pressure) etc. are recorded.

As with the SPRAYER the WINDOX software allows for integrated or time resolved measurements with up to 2000 particles size distributions per second.

INHALER
Aerosol dispersion for laser diffraction
0.25 µm - 1750 µm

Central Unit

INHALER central stage completely equipped

Central stage with complete set-up

Rear View

Rear view of the INHALER

Rear view of the INHALER with Vacuum Control Unit and Fine Particle Collector

Dry Powder Inhaler (DPI) Application

INHALER with eleongation front cylinder and Ventury meter

INHALER Central Unit with DPI, Elongation Cylinder and Venturi Meter

INHALER with pre-separator and 90° elbow

INHALER Central Unit with DPI, 90° -Elbow, Pre-separator and Venturi Meter

Control Panel

The INHALER can be operated via software control or manually via the control panel.

INHALER control panel

The control panel allows for the set-up of all parameters, the flow adjustment and the control of the vacuum system.

Add this page to my favorites Send this page to my colleague Print this page