sensor: |
OPUS | fingerprobe diameter = 89 mm length = 330 - 3500 mm for pipes from DN 10, vessels,... |
| measuring range: | particle size solids concentration media |
0.01 ?m - 3000 ?m 31 classes, xmax/xmin < 1000 1 - 70 Vol % typ. suspensions, emulsions |
| principle: | ultrasound | 100 kHz to 200 MHz dynamics fixed: 150 dB precision: 0.1dB typ. |
| evaluation: | extinction function | library > 300 products |
| operation: | process control software |
optional via PLC WINDOWS XP Prof / 2000 |
| performance: | repeatability comparability x10, x50, x90 measuring time |
SD. < 0.5% (repeated sample) SD. < 1.0% (riffled sample) SD. < 5% mean rel. SD < 1 minute, typ. |
| environment: | temperature (product) pressure protection class materials in contact with media |
≤120°C (standard) or 150° C (as option) ≤ 40 bar IP65 or ATEX adaptable to application stainless steel, Sigradur?, Teflon?, |