Technical specfications

principle: suspension layer 1- 10 mm
dosing: fluid circulation
volume
pumped & stirred
500 ml - 1 l
operation: software-controlled fast filling & draining
temperature control
application: temperature
material in contact with media
≤ 80°C
stainless steel, Sigradur?,
fibre enforced Tesflon?
sensor: NIMBUS lab system
measuring range: particle size

solids concentration
media
0.01 ?m - 3000 ?m
31 classes, xmax/xmin < 1000
1 - 70 Vol % typ. (viscosity dependent)
suspensions, emulsions
principle: ultrasound 100 kHz to 200 MHz
fixed dynamics: 150 dB
precision: 0.1 dB typ.
evaluation: extinction function evaluation Software
WINDOX KSIGMA
operation: sotware controlled standard operating procedures
performance: repeatability

comparability 
x10, x50, x90
measuring time
s. < 0.5% (repeated sample)
s < 1.0% (riffled sample)
s. < 5% mean rel. SD

< 1 minute, typ.

more ...

NIMBUS
Particle size analysis with ultrasound 0.01 ?m - 3000 ?m
in suspensions and emulsions of high concentration

symbol: wet dispersion for suspensions or emulsions
お気に入りに追加 このページをメール送る このページを印刷