Technical Specifications

sample coupler: customer screw sampler, etc ....
principle: fully adaptable to cutomer's existing sampler highest flexibility
analysis system: MYTOS 0.25 ?m - 1750 ?m
dispersion
principle:
RODOS-aerosol jet particle-particle-collisions
particle-wall-collisions
velocity gradients
analysis
principle:
HELOS laser diffraction λ = 632.8 nm
measurement: multi-element-detector
auto-alignment
measuring frequency
31 semi-circular elements
permanent
2000/s
evaluation: Fraunhofer MIE as an option
measuring range: optical modules R2 - R6
operation: at-line

RS232 / RS484-interface
optional via PLC
Windows 2000
performance: repeatability
comparability 
x10, x50, x90
SD. < 0.3 % (riffled sample)
SD. < 1 % mean rel. SD
/Δx/ < 2.5 % rel. max. SD.
environment: temperature (product)
pressure
protection class
optional
materials in contact
with media
sample disposal
max. 60°C
max. 10 bar
IP55 (control box)
Ex
Stainless Steel or customised

back into process or external
QS-system: certification
validation

 

standard test procedure
FDA approved